0.5 V output at the input of 100 x 10-6 in strain
(With bridge power supply at DC2 V)
Output
MONITOR ±10 V, ±5 mA (Load resistance 2 k-ohm or more)
OUTPUT ±10 V, ±30 mA (Load resistance 330 ohm or more)
Output resistance 10 ohm or less
Output capacitive load 0.1 micro-F or less
Non-linearity
±0.01 % F.S.
Standard equivalent strain(CAL)
Set with digital switch
±1 to ±9 999x10-6 in strain,
Accuracy ±0.2 %F.S. or less at each point
Sensitivity adjustable range(ATT)
1, 1/2, 1/5, 1/10, 1/20, 1/50, 1/100, 0
Accuracy ±0.2 %F.S. at each point
GAIN volumu valiable range
1 to 1/2.5 times continuous variable (VAR)
Frequency response range
DC to 200 kHz +1 dB, -3 dB (Filter at W/B)
Low-pass filter
10 Hz, 100 Hz, 1 kHz, 10 kHz, W/B Third bessel type
S/N ratio
46 dBp-p(ATT=1) with W/B condition
55 dBp-p with 10 Hz to 10 kHz
Display
Output voltage display 0.00 to ±10.00
Digital display (Green LED)
Sampling Approx.4 times/s
0.00 or -0.00 flashes when output voltage is in non-linear zone
Temperature efficient
Zero shift ±0.1x10-6 in strain / 'C
Sensitivity variation ±0.01 %F.S./ 'C
Efficient due to time
Zero shift ±5x10-6 in strain / 8 h
Sensitivity variation ±0.05 %F.S./8 h
Efficient due to power supply
Zero shift ±1x10-6 in strain / variation of power supply voltage within ±10 %
Sensitivity valliation ±0.01 %F.S.
/variation of power supply voltage within±10 %
Operating temp./humidity range
Temperature :-10'C to 50'C, Humidity 85 %RHor less (Non condensing)
Power supply voltage
AC100 V±10 V, DC12 V 50/60 Hz
Power consumption
Approx. 10 VA(at AC100 V), Options DC12 V Approx. 0.8 A
Outline dimensions(WxHxD)
42 mmx176 mmx284.4 mm(excludes protruding parts)
Weight
Approx. 1.5 kg
Dynamic strain amplifier DSA - 631
Feature
Eqquipped with wire length compensating function and open circuit check function.
High sensitivity and high gain
Excellent visibility and operation
Dual power supply types
Specifications
Specifications
Measuring point
1 point per 1 unit
Applicable gage resistance
60 ohm to 1k-ohm
Gage factor
2.00
Bridge power supply
AC0.5 V, 2 V, 5 kHz sine wave, Sync input/output signal 2.5 VAC
Balance adjusting range
Deviation of resistive value : Approx. ±2 % (±10 000x10-6 in strain)
Capacitive imbalance : Approx. 2 000 pF
Balance adjusting Accuracy ±0.4x10-6 in strain (RANGE=200, without FINE, BV=2 V)
Maximum input range
±200 000 x 10-6 in strain (RANGE=20k, FINE=x2.5, BV=0.5 V)
Voltage sensitivity
±10 V within ±200 x 10-6 in strain
Shift of measuring range
200, 500, 1k, 2k, 5k, 10k, 20k (x10-6 in strain, x2/BV in value), OFF
Fine adjustment continuously changeable in FINE ~ RANGE
2 steps changing amount can be selected.
Non-linearity
±0.1 %F.S.
Standard equivalent strain(CAL)
Set by digi-switch : ±1 to 9 999 x10-6 in strain
Accuracy ±0.5 %rdg+0.5x10-6 in strain
Zero drift : Within ±0.1x10-6 in strain /'C
Sensitivity : Within ±0.05 %F.S./'C
Efficient due to time
Zero drift : Within ±0.5x10-6 in strain /24 h
Sensitivity : Within ±0.2 %F.S./24 h
Noise
2.0 x10-6 in strain p-p RTI
(W/B, RANGE=200, Without FINE, BV=2 V, 120 ohm in bridge)
0.6 x 10-6 in strain p-p RTI
(DC=100 Hz, RANGE=200, Without FINE, BV=2 V, 120 ohm in bridge)
Output
OUTPUT1 ±10 V, ±5 mA
OUTPUT2 ±10 V, ±10 mA
Output impedance 0.5 Ω or less,
Capacitive load Operable up to 0.1 μF
Output adjustment
OUTPUT2 ADJ
(Can be independently varied continuously from 1 to 1/10)
Operating temp./humidity range
Temp. -10'C to 50'C,
Humidity 20 % to 85 %RH or less(Non condensing)
Power supply voltage, Power consumption
AC85 V to AC132 V/AC180 V to AC264 V
(Internal switch must be changed) 9 VA or less
DC10 V to DC30 V, 6 VA or less
Outline dimensions
(W x H x D)
49.5 mm x 143 mm x 253 mm (excludes protruding parts)
Weight
Approx. 1.35 kg
Precision Calibrator PSC-350
Feature
350 ohm precision calibrator
Output calibrated by connecting this unit with a digital indicator, etc., for the strain gage based transducer.
Specifications
Specifications
Rated output
0.5 mV/V range 10 steps
2 mV/V range 10 steps
5 mV/V range 10 steps
10 mV/V range 10 steps
Error
0.02 %F.S. (Temperature range from 15'C to 25'C)
Excitation, maximum
DC 12V
Zero balance
±0.001 mV/V
Input resistance
350Ω ±7Ω
Output resistance
350Ω ±7Ω
Temp. effect on zero
0.03 %F.S. / 10'C
Temp. effect on output
0.03 %F.S. / 10'C
Operating temp./humidity range
Temperature 0'C to 50'C, Humidity85%RH or less(Non condensing)
Outline dimensions(WxHxD)
185mm x 47mm x 125mm (excludes protruding parts)
Weight
Approx. 1.3kg
Static Strain Digital Indicator PSD-704
Feature
Indicator for strain gage or strain gage
applied transducer.
Check strain gage based transducers in
mV/V.
Measure I/O resistance of strain gage based transducers. Insulation resistance measurement between load cell body and each cable wire is also available.
Applied RS-232C interface as a standard.
Specifications
Specifications
Bridge power supply
DC 2V±0.02 V Within 35 mA
Target for measurement
Strain measurement, Measurement on output of strain gage applied transducer, Meadurement on resistance value.
Measuring method
Deflection method
(under measuring on strain or output of strain gage application)
Temperature efficient
Zero point ±0.2 x 10-6 in strain/'C, Sensitivity ±0.01 %F.S./'C
(After 15 min. of warming up time with the range of x1)
Efficient due to time
Zero point ±0.2 x 10-6 in strain/8 h, Sensitivity ±0.01 %F.S./8 h
(Temperature variation within ±2'C with the range of x1)
A/D sampling
4 times/s
Display
Dot matrix type liquid crystal display (with back light LED type)